Resuscitation With Master Flux Plus - Tecno-gaz Master Flux Plus Instrucciones De Uso

Ocultar thumbs Ver también para Master Flux Plus:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 133
7.3

RESUSCITATION WITH MASTER FLUX PLUS

RESUSCITATION procedures WITH MASTER
FLUX PLUS must be carried out with the %N2O
knob that regulates the percentage of nitrous oxide
in position 0%.
SLOWLY open the oxygen cylinder valve (turn the valve counter-clockwise); make
sure that the nitrous oxide cylinder is closed (turn the valve clockwise).
Take the MASTER FLUX PLUS mask kit off by sliding it off of the spiral tube.
Insert the relative fitting (30) into the spiral tube.
Remove the bag (F) and put the cap (T).
Connect the end of the oxygen treatment tube (31) to the fitting (30).
Connect the other end of the oxygen treatment tube to the bottom stud of the resuscitation
bag (32) through the relative coupling.
Connect the mask (33), to the valve on the resuscitation bag.
Turn the FLOW knob to deliver Oxygen at the required flow rate, CHECK THE FLOWS
DISPLAYED ON THE FLOW COLUMNS.
The nitrous oxide flow meter =0
The oxygen flow meter = SET FLOW
Place the mask on the patient.
Act on the resuscitation bag in accordance with the resuscitation instructions.
Once use of the device is finished it is IMPORTANT to close the valve on the Oxygen
cylinder (by turning it clockwise).
*(KIT Art. 1504/A Not included – see figure 19)
ENGLISH
27

Publicidad

Tabla de contenido
loading

Tabla de contenido