ALFUNCTIONS
6.1

Malfunctions

Component
Vaco 200 unit
Vacuum
electrodes
Sponges
6.1.1
Equipment cannot be switched on
1. Check if the mains voltage has failed.
2. Check if the main switch is switched on ("I").
3. Check if the power cord and the fuses are in order. If necessary,
replace the fuses. See §6.1.2.
4. Contact your dealer if the equipment still cannot be switched on.
6.1.2

Replacing the fuses

1. Switch the main switch off ("O").
2. Unplug the power cord from the equipment.
3. Pull the fuse holder carefully out of the equipment. If necessary, use
a screwdriver.
4. Replace both fuses. If necessary, order new fuses from your dealer.
5. Install the fuse holder and plug in the power cord.
6. Switch the main switch on again ("I").
6.1.3
Equipment does not react to commands
The safety system of the equipment has ascertained a fault. You cannot
continue to work.
1. Disconnect the connection to the patient.
2. Switch the main switch off ("O").
3. Wait 5 seconds and switch the main switch on again ("I").
4. Contact your dealer if the fault reappears.
Vaco 200
,
SERVICE AND GUARANTEE
Problem
Equipment cannot be
switched on
Equipment does not react to
commands
Contamination by ionization See §6.1.4.
Furring
Bad conduction
Solution
See §6.1.1.
See §6.1.3.
Replace the sponges
Replace the sponges