Electromagnetic compatibility
14.3.2 Guidelines and manufacturer's
declarations - electromagnetic interference
immunity, non-life-supporting equipment
The device is intended for use in the electromagnetic
environment specified below. The customer or user
of the device should ensure that it is operated in
such an environment.
The environment is described as follows:
Interference immunity
IEC 60601 test level / compliance level
− Electromagnetic environment – guidelines
Conducted HF disturbance variables according to
IEC 61000-4-6
6 V
eff
150 kHz - 80 MHz
Radiated HF disturbance variables according to
IEC 61000-4-3
3 V/m
80 MHz - 2.7 GHz
− Portable and mobile radios should not be used
at a distance from the product, including cables,
that is less than the recommended safety distance
(30 cm) calculated according to the equation
applicable to the transmission frequency.
− The field strength of stationary radio transmitters
should be lower than the compliance level at all
frequencies according to on-site testing.
− Interference may occur in the vicinity of
equipment marked with the following symbol:
Remark:
At 80 MHz and 800 MHz the higher frequency range
applies.
96200-114 – 10 – 15.07.2020
14.4 Recommended safety distances
Guidelines and manufacturer's declaration / high-
frequency wireless communication devices
385
TETRA 400 (380...390 Mhz) 1.8
GMRS 460, FRS 460
450
(430...470MHz)
710,
LTE Band 13/ 17
745,
(704...787MHz)
780
GSM 800/900, TETRA 800,
810,
iDEN 820, CDMA 850, LTE
870,
Band 5
930
(800...960MHz)
GSM 1800, CDMA 1900,
1720,
GSM 1900, DECT, LTE Band
1845,
1/ 3/ 4/ 25, UMTS
1970
(1.7...1.99GHz)
Bluetooth, WLAN,
802.11 b/g/n, RFID 2450,
2450
LTE Band7
(2.4...2.57GHz)
5240,
WLAN 802.11 a/ n
5500,
(5.1...5.8GHz)
5785
0.3
27
5
0.3
28
0.2
0.3
9
2
0.3
28
2
0.3
28
2
0.3
28
0.2
0.3
9
57