10.0 Technical data / maintenance / calibration
Measuring range [N]
Resolution [N]
Measuring frequency
Display update time
AC-Charger
User manual
Attachments (FMI-901 M4 / FMI-901 M6 /
FMI-901 M10)
Accuracy (F.S. +/- last significant digit)
Temperature drift (Tk relative) per K
Zero-point-drift (Tk absolute) per K
Tracking (Tk suppressed)
Overload signal
RS 232 C interface
100Hz | Baud rate settable
USB 100Hz
Analogue signal -10V...0...10V)
Memory function
Threshold value function (Limit)
Temperature range (operation)
Temperature range (storage)
Protection Code
Weight
Size (L x W x H)
The gauges, apart from expendable and wearing parts, are maintenance-free. Depending on the area and
frequency of use, gauges should be regularly, namely at least once a year, be checked and calibrated. Our
calibration service (e-mail: service@alluris.de) includes a technical check of the instrument, calibration (factory
calibration) and, if necessary, adjustment of the measurement range.
FMI-100
FMI-200
FMI-210
0-2
0-20
0-5
0-50
0-20
0-20
0-200
0-50
0-50
0-500
0-200
0-200
0-500
0-500
0,001
0,001
0,01
0,001
0,001
0,01
0,01
0,01
0,1
0,01
0,01
0,1
0,1
0,1
(-/ /-)
( / /-)
( / /-)
+/-
+/-
0,2%
0,05%
0,2%
+/-
+/-
0,02%
0,02%
0,02%
+/-
+/-
0,02%
0,02%
0,02%
450 g
147 x 75 x 38 mm
32
FMI-220
FMI-230
0-2
0-2
0-2
0-5
0-5
0-5
0-20
0-20
0-50
0-50
0-200
0-200
0-500
0-500
0-1000
0,001
0,001
0,001
0,001
0,01
0,01
0,01
0,01
0,1
0,1
0,1
0,1
0,1
0,1
1
1000 Hz
1-20 Hz (selectable)
( / /-)
( / /-)
+/-
+/-
+/-
0,05%
0,2%
+/-
+/-
+/-
0,02%
0,02%
+/-
+/-
+/-
0,01%
0,01%
automatic compensation
0° ... 40° (max.. 85%rF)
- 5° ... 45 (dry storage)
IP 40
FMI-300
FMI-400
0-1000
0-1000
0-2500
0-2500
0-5000
1
1
1
1
1
(-/-/ )
(-/-/ )
+/-
+/-
0,2%
0,2%
+/-
+/-
0,02%
0,02%
+/-
+/-
0,02%
0,02%
900 g
214 x 82 x 55 mm