2. 2 REPLACEMENT
The P
LED handpiece must be replaced when
IEZOTOME
the thread no longer holds the tips adequately and
when repeated sterilization cycles lead to visible
damage of the casing (cracks, scratches).
The optical guide and the light ring of the P
LED handpiece must be replaced when the light
intensity is no longer sufficient and when repeated
sterilization cycles lead to visible damage (cracks,
scratches).
2. 3 TROUBLESHOOTING
Refer to the table
F
AULT
No spray
Incorrect irrigation line connection
Obstructed tip
No water at tip but
vibration present
Incorrect spray adjustment
Insufficient power -
Worn or deformed tip
Low vibration
Incorrect tip tightening
Defective connector contacts of the
No ultrasonic output
P
Cut wire in handpiece cord
LED ring not fitted to the handpiece
Faulty LED ring
Faulty LED ring connector contacts
No light
LED ring poles reversed
Faulty handpiece and/or cord connector
contacts
Other
P
C
OSSIBLE
AUSE
LED
handpiece-cord assembly
IEZOTOME
III - LIABILITY
The manufacturer's liability shall not be pledged in
the following cases:
- Maintenance or repair by persons not authorized by
the manufacturer.
IEZOTOME
- Use on an electrical installation that does not
comply with the regulations in force.
- Other uses as those specified in this manual.
The manufacturer reserves the right to modify the
handpiece and/or the user's manual without prior
notice.
Check the irrigation line connection
Remove obstruction
Adjust spray
Replace tip
Tighten with wrench
Clean cord connector and socket unit contacts
Return to customer service for P
cord assembly replacement
Fit the LED ring
Replace the LED ring
Clean the LED ring connector contacts
Position the LED ring according to the "+" foolproofing
marks.
Clean the handpiece and/or cord connector contacts
Return to SATELEC after-sales
6
R
EMEDY
LED
handpiece-
IEZOTOME