Turn the C 1 calorimeter off and on again; at this point, the hea-
ter will be recognized by the C 1 calorimeter.
The error message "Heater not
NOTICE
found" should not be displayed
(see "Error codes" section).
Open the water tap slowly.
Check connections for tightness.
CAUTION
Check the cooling water temperature on the C 1 calorimeter in
the "Accessories" menu under the "Condenser test" item.
Then choose the corresponding measuring mode for the C 1
calorimeter in the "settings" menu (see "Use for the intended pur-
poses" section).
If the water inlet temperature is
NOTICE
too high, select a higher tempe-
rature range for the C 1 to allow
a measurement to be performed.
If the water inlet temperature is
NOTICE
low, you should always select the
lower operating temperature for
the C 1 to save time and energy.
Proceed as follows in the case of error code:
Switch the device off,
Perform corrective measures as described below,
Perform a restart.
Error
Cause
Heater not
C 1.20 heater is not recognized by
found
the C 1 calorimeter.
Perform a system test in accordance with the C 1 calorimeter
operating instructions or a measurement.
Error Codes
Effect
No heater
If the actions described fail to resolve the fault or another error
code is displayed then take one of the following steps:
- Contact the service department,
- Send the device for repair, including a short description of the
fault..
Fig. 20
Solution
Always turn the heater on first.
Check the heating connection cable plug contact.
Switch the C 1 calorimeter off and on again.
Fig. 18
Fig. 19
16