Programme Sequence; Premature Programme Abort; Cleaning And Maintenance - Dreve EyeVolution Instrucciones De Uso

Foto-polimerización basado en la luz de leds
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

EyeVolution

7.3 Programme sequence

Important: In order to protect
your eyes, polymerisation is only
possible when the device cover
is closed.
The light used is not suitable for direct
eye contact and can cause retinal dam-
age over the long term when unfiltered.
The cover features a light-filtering
inspection window (13) to facilitate
visual inspection of the polymerisation
process.
An on-going time programme is always
interrupted by opening the cover (12);

7.4 Premature programme abort

An on-going time programme can be
ended and deleted at any time by press-

8. Cleaning and maintenance

Always remove the mains plug from the
mains power supply before carrying out
any cleaning or maintenance work.
Ideally, the device is cleaned using a
soft cloth, possibly a slightly dampened
sponge and a mild cleaning agent.
24
®
the LED assigned to the respective
programme flashes. The time remain-
ing is saved in the electronics module
and continues once the polymerisation
chamber is closed again.
Once a time programme is complete,
2 brief signal tones sound and the
polymerisation LEDs and rotary motor
switch off automatically. The green
LED indicating the programme on the
operating foil (3) is also deactivated. A
ring of 6 white LEDs stays active and
is only dimmed down after the device
cover is opened.
ing the activated programme key.
Water and cleaning agent must not
penetrate the device.
The LED circuit board (9) must not
come into contact with damp materi-
als. Minor soiling can be removed using
a soft brush.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

D-38100

Tabla de contenido