HEINE EN200/EN200-1
These instructions apply to the following products of the HEINE Wall Transformer series:
EN200 and EN200-1
Please read and follow these instructions for use and keep them for future reference.
Intended Use
The HEINE EN200 wall transformer is a dual-channel, dimmable, 3.5 V voltage supply unit with an
additional 5 V output for medical diagnostic equipment from HEINE.
The HEINE EN200-1 wall transformer is a single-channel, dimmable, 3.5 V voltage supply unit for
medical diagnostic equipment from HEINE. The product can be used either in combination with the
EN200 wall transformer or as a stand-alone device.
It should only be operated by qualified medical personnel within a professional healthcare facility.
For U.S. only:
Federal law restricts this device to sale by or on the order of a Physician or Practitioner!
Warnings and Safety Information
CAUTION! Indicates potential hazardous situations. Ignoring the corresponding instructions may
lead to dangerous situations of mild to moderate extent. (Background color yellow; foreground
color black).
NOTE! This symbol indicates valuable advice in terms of set up, operation or maintenance,
as applicable. Notes are important, but not related to hazardous situations.
Assembly and disassembly
Assembly accessories
Allen wrench 1.3
To ensure all-pole disconnection from the mains at all times, the ME device must be installed so
that the switch mode power supply is accessible and disconnectable.
Before mounting the device, check whether you need to use a special wall plug for the subsurface
of your wall and whether the subsurface exhibits sufficient load-bearing capacity for the meachnical
stress.
The wall plugs supplied with the product are universal wall plugs and suitable for most building
materials (e.g. concrete, solid brick, brick).
1. Hold the drill template horizontally in place on the desired position and mark the drill holes. Drill a
hole of at least 40 mm deep using a Ø 6 mm drill.
2. Fit wall plugs.
3. Use a suitable screwdriver to tighten the two upper screws with a distance of approx. 3 mm to the wall.
4. Hang the device – disconnected from the mains – on the two screw heads, press firmly and push
downwards. Check to make sure that the device is correctly positioned on both screw heads.
5. Screw in the two lower screws.
To disassemble, disconnect the device from the mains first, loosen the two lower screws, push upwards
and take off the wall.
USB connector
On the back of the EN200 is a USB-A output for the supply of voltage to other medical diagnostic
equipment from HEINE.
The EN200-1 has a micro-USB input on the back for the supply of voltage via a medically approved
USB power supply or via the EN200.
Connect the HEINE USB cord and run the cable via the housing canal to the outside on the left or right
hand side. To do so, press the cable into the clamping hubs, which act as strain relief.
6/44
MED 113066
2019-02-06
ENGLISH