Program 3: Low Frequency
The Low Frequency program utilizes the same parameters of Stage 2 of the WiTouch Pro Exclusive
treatment program.
Mode
Pulse Duration (sec) *Measured positive phase
Pulse repetition frequency (Hz)
Output voltage (V) (Maximum)
Load Range that is valid for these parameters (Ω)
Effect of load impedance on these parameters
Program 4: Hi-Lo Combo
The High-Low Combo program alternates between high frequency stimulation and low frequency
stimulation.
Note:
This program is only accessible with the WiTouch Pro app
Mode
Pulse Duration (sec) *Measured positive phase
Pulse repetition frequency (Hz)
Output voltage (V) (Maximum)
Load Range that is valid for these parameters (Ω)
Effect of load impedance on these parameters
Timer Control: (mins) 30
Power Supply: Two (2) AAA batteries (Internally Powered)
Size (D x W x H): 0.7" x 7.5" x 3.5" (18 mm x 191 mm x 90 mm)
Weight (including battery): 4.8 oz (136 g)
Battery Life: Over 100 treatments based on typical use
The expected service life for the WiTouch Pro device is 2 years.
Shelf life of gel pads is two years (see "Use By" date printed on pouch).
Safety Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, CAN/CSA C22.2
No. 601.1,
Made in USA
Troubleshooting
Event
Low Battery
Device transition from OFF to ON
Device transition from ON to OFF
Device transition from ON to
start treatment
Note:
This program is only accessible with the WiTouch Pro app
External load resistance can be up to 10kΩ.
If this load resistance is exceeded, the
nominal current flowing to the patient may
be less than the set target stimulation current.
Stage 1
External load resistance can be up to 10kΩ.
If this load resistance is exceeded, the
nominal current flowing to the patient may
be less than the set target stimulation current.
LED Color
Yellow
Slow flashing during standby when the
device is ON. Rapid flashing during the
stimulation treatment.
Green
Slow flashing during standby when the
device is ON.
None
Green flashing stops.
Green
Rapid flashing during treatment.
Stage 1
240μs
5 to 10Hz
(Variable)
55V
500Ω
250μs
2Hz
55V
500Ω
LED Sequence
Stage 2
150μs
70Hz
55V
21