Program Descriptions - Hollywog WiTouch Pro Manual Del Usuario

Ocultar thumbs Ver también para WiTouch Pro:
Tabla de contenido

Publicidad

Program Descriptions

Program 1: WiTouch Pro (WT Pro) Exclusive
The WiTouch Pro exclusive, patented 3-stage waveform incorporates both clinical theories of
TENS to provide pain relief. The 30-minute stimulation treatment is delivered as follows:
Stage 1: 5 minutes of high frequency stimulation (sometimes described as a strong tingling
sensation) that can initiate a feeling of pain relief by suppressing the transmission of pain signals
in nerves. This stage provides a high sensory sensation and allows you to establish a comfortable
intensity setting for the entire treatment.
Stage 2: 20 minutes of low frequency stimulation (sometimes described as a gentle tapping sensation)
that can initiate an increased endorphin release in the body to reduce the sensitivity to pain for an
extended period of time following the 30-minute treatment.
Stage 3: 5 minutes of high frequency stimulation, providing the same high sensory sensation
experienced in Stage 1. This stage can allow you to maintain the feeling of pain relief and
complete the overall treatment with a comfortable high sensory sensation.
Mode
Pulse Duration (sec) *Measured positive phase
Pulse repetition frequency (Hz)
Output voltage (V) (Maximum)
Load Range that is valid for these parameters (Ω)
Effect of load impedance on these parameters
Program 2: High Frequency
The High Frequency program utilizes the same parameters of Stages 1 and 3 of the WiTouch Pro
Exclusive treatment program.
Mode
Pulse Duration (sec) *Measured positive phase
Pulse repetition frequency (Hz)
Output voltage (V) (Maximum)
Load Range that is valid for these parameters (Ω)
Effect of load impedance on these parameters
Program 3: Low Frequency
The Low Frequency program utilizes the same parameters of Stage 2 of the WiTouch Pro Exclusive
treatment program.
Mode
Pulse Duration (sec) *Measured positive phase
Pulse repetition frequency (Hz)
Output voltage (V) (Maximum)
Load Range that is valid for these parameters (Ω)
Effect of load impedance on these parameters
Note:
This program is only accessible with the WiTouch Pro app
Note:
This program is only accessible with the WiTouch Pro app
Stage 1
Stage 2
120μs
240μs
80 to 120Hz
5 to 10Hz
(Variable)
(Variable)
55V
500Ω
External load resistance can be up to 10kΩ.
If this load resistance is exceeded, the
nominal current flowing to the patient may
be less than the set target stimulation current.
Stage 1
120μs
80 to 120Hz
(Variable)
500Ω
External load resistance can be up to 10kΩ.
If this load resistance is exceeded, the
nominal current flowing to the patient may
be less than the set target stimulation current.
Stage 1
240μs
5 to 10Hz
(Variable)
500Ω
External load resistance can be up to 10kΩ.
If this load resistance is exceeded, the
nominal current flowing to the patient may
be less than the set target stimulation current.
Stage 3
120μs
80 to 120Hz
(Variable)
55V
55V
55V
55V
9