Hollywog WiTouch Pro Manual Del Usuario página 8

Ocultar thumbs Ver también para WiTouch Pro:
Tabla de contenido

Publicidad

Using the WiTouch Pro Device in Manual Mode
The WiTouch Pro device can be operated without the WiTouch Pro app by using the buttons on the device.
Note:
When using the WiTouch Pro device with the buttons on the device, the default treatment program will
be program 1 (WiTouch Pro [WT Pro] Exclusive).
1. Follow the steps under the section Conducting a Treatment and apply the WiTouch Pro device to
your back as instructed.
2. Press the (+) button to start the WiTouch Pro Exclusive treatment program. Press the (+) button
again, one press at a time, to increase the intensity of the stimulation until it is at a comfortable
level. Press the (-) button to decrease the intensity of stimulation as desired.
3. The treatment will automatically stop after 30 minutes. To stop the treatment at any time
during treatment, press the ON/OFF one (1) time. This stops the treatment and the WiTouch Pro
device will remain in standby mode to run additional treatments.
in standby mode for 2 hours. After 2 hours of non-use, the device will automatically shut off to
save battery life.
Removing the WiTouch Pro Device
IMPORTANT! Do not remove the device until the treatment has stopped.
1. After treatment, or when you want to remove the device, grasp the edge of the device and gel pad to
ensure the gel pad does not stay on the skin. Slowly peel the device away from the skin.
2. Align and place the green protective liners back on the gel pads. Ensure the pads are completely covered.
Battery Replacement
When battery replacement is needed the LED will flash as follows:
• LED flashes yellow every two (2) seconds when the unit
is On and in Standby
• LED flashes yellow rapidly during treatment
Note:
Fully depleted batteries will have no flashing LED.
To replace the device batteries:
1. Using the included #0 Phillips head screwdriver,
remove the cover.
2. Remove the old batteries, and place the new batteries in the correct (+) and (-) polarity.
3. Replace the battery compartment cover.
Note:
Consult your local authorities for proper disposal of batteries, device, and accessories.
Maintenance and Storage
• Use a damp cloth and mild soap to gently wipe clean after each use, when the is device is
soiled, or to remove build-up of gel-pad residue. Remove any soap residue from cleaning.
Damage to the device and/or accessories will occur if submerged in water or other liquids.
WiTouch Pro device is manufactured with water detection technology to detect
exposure to water that may cause damage to the electronics and void the product warranty.
• The unit should be routinely checked before each use to determine all controls function normally.
• The device should be operated, transported and stored at temperatures between 50° F and
104° F (10° C and 40° C), with relative humidity between 30% - 85%.
• Store the device, and accessories in the original packaging when not in use.
• Remove batteries before extended storage to reduce likelihood of battery corrosion or leakage.
• If the device is not working properly, stop using immediately. Do not disassemble or modify the device.
Contact Hollywog Customer Service at: Phone: +1 423 305 7778; or Email: info@hollywog.com.
• To purchase replacement gel pads for your device, you can find them wherever the device is sold.
• If you have questions, please call Hollywog Customer Service at: Phone: +1 423 305 7778; or
Email: info@hollywog.com.
Technical Specifications
Channels: Single channel
Waveform: Asymmetric biphasic square pulse
Pulse Amplitude (Maximum): 110 mA (55 volts) at 500 ohm load
Pulse Frequency: (Hz) 2-120
Pulse Duration: (µs) 120-250
Note: