6 MAINTENANCE

Lubricating the seals (once a year)

-
Remove the probe from the tank by disconnecting the power
cable, sliding fork F1 out and removing the complete probe
(pulling upwards).
-
Slide out the fork F2 and remove the cover C2
-
Lubricate the seals G1 and G2 with silicon grease.

Anomalies and solutions

TYPE OF ANOMALY
The display does not come
A
on
The digital display shows
B
an error message "C Err".
The digital display shows an
C
error message "Probe nc".
The digital display shown an
D
error message "Cap Err".
The display indicates the
E
level only up to a certain
point.
The display indicates the
F
level only from a certain
point onwards.
G
Not all the LEDs come on.
H
The LEDs flash.
In the event of any other errors not described in the above table, please consult the Geoline assistance service.
30
CAUSE
The power cable is not delive-
ring electricity.
The fuse has blown.
Probe supply cable connec-
tion error.
Incorrect power voltage.
No probe calibration pro-
gramme.
Probe clogged (scale or solid
residues).
Incorrect probe calibration.
Probe power cable connection
error.
Incorrect power voltage.
No probe calibration pro-
gramme.
U s e
&
M a i n t e n a n c e
SOLUTION
Check that the cable is correctly connected; repla-
ce the cable and/or the display if necessary.
Replace the fuse.
Check the probe power cable is connected
correctly, tighten the plug head fixing screws if
required.
Check the condition of the 3 tabs on the probe
head, and remove any scale. Having done this,
switch the display off and then on again.
Check that the voltage matched the minimum and
maximum values foreseen in this manual.
Enter the calibration programme in the electronic
card (consult an authorised Geoline retailer).
Remove the probe and clean carefully.
Consult the manufacturer.
Check the probe power cable is connected correctly,
tighten the plug head fixing screws if required.
Check the condition of the 3 tabs on the probe
head, and remove any scale. Having done this,
switch the display off and then on again.
Check that the voltage matches the minimum and
maximum values foreseen in this manual.
Enter the calibration programme in the electronic
card (consult an authorised Geoline retailer).
M a n u a l
6
6
ENG