Manual URYXXON
The strip is illuminated with an LED and a detector registers the intensity of light
reflected by the test strip at three different wavelengths. Using an internal cali-
bration, the results are calculated from the reflection values. Whenever samples
are strongly alkaline, a density correction is automatically conducted.
2.5.2. Functional principle URYXXON
A measurement is started by placing a strip on the holder. If the Autostart-feature
is turned off, the measurement is started by pressing the start panel
the display. The result is shown on the display, printed out and released via the
interfaces after the measurement has been completed. After three minutes the
instrument will go to stand-by. Touching the screen will reactivate the instrument.
All user inputs are performed via the touch-screen (Chapter 3.11 "Use of the
instrument").
2.6. Safety warnings
The following safety warnings are used throughout the manual:
- DANGER: defines a danger which can result in serious injury or death.
- ATTENTION: defines a possible danger which can result in simple - mild injury.
This symbol is also used to indicate user errors which can result in malfunction
or damage to the device.
- NOTE: provides necessary additional information.
68
®
Relax
Pic. 5: Measuring Principle
®
Relax
Manual URYXXON
®
Relax EN, V6.00 / 11.20