Specifications - Masterflex C/L 77120-32 Manual De Operación

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

SPECIFICATIONS

Output:
Operating Speed:
Models 77120-32,
and 77120-52
Models 77120-42,
and 77120-62
Maximum No. of Tubes:
Direction of rotation:
Input:
Supply Voltage/Frequency:
Models 77120-32
Models 77120-42
Models 77120-52 and 77120-62
Power Input:
For 6 and 60 rpm models
Installation Category:
77120-32, -42
77120-52, -62
Remote Start/Stop:
8
®
1 to 6 rpm
10 to 60 rpm
2
Clockwise and
Counterclockwise
115V AC nominal, 50/60 Hz
(90–130V AC) @110 mA AC
230V AC nominal, 50/60 Hz
(190–260V AC) @75 mA AC
115V AC nominal, 50/60 Hz
(90–130V AC) @175 mA AC
230V AC nominal, 50/60 Hz
(190–260V AC) @108 mA AC
12V DC nominal
(11.5–16.0V DC) @ 1.0 A DC
11.5–16.0V DC to terminals
1 and 2 of terminal strip or
External Power Supply
Category II per IEC664
(Local level—appliances,
portable equipment, etc.)
Category I per IEC664
(Signal Level)
Contact Closure connection
at terminal strip contacts 3
and 4
Construction:
Dimensions (L × W × H):
Weight:
Color:
Material:
Pump Head:
Case:
Enclosure Rating:
Environment:
Operating Temperature:
Storage Temperature:
Humidity (Non-cond):
Altitude:
Pollution Degree:
Noise level:
Compliance (For CE Mark):
EN61326-1/A2: 2001 (EMC Directive)
EN809 (Machinery Directive)
5.3 in × 3.5 in × 3.5 in
(13.5 cm × 8.9 cm × 8.9 cm)
2 pounds (0.91 kg)
Black
PPS, polypropylene
and stainless steel
NORYL
®
IP22 per IEC529
32°F to 105°F (0°C to 40°C)
–49°F to 149°F (–45°C to 65°C)
10% to 90%
Less than 6600 ft (2000 m)
Pollution Degree 2 per IEC664
(Indoor usage — lab, office)
<70dBa @ 1 meter

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

C/l 77120-42C/l 77120-52C/l 77120-62

Tabla de contenido