CALIBRATION
Calibration
This scale has been factory calibrated, and does not require calibration prior to use. If
required, the scale can be calibrated. Calibration of your 498KL scale is performed using
kilograms (KG) or pounds (LB), according to the units of measure used upon entering into the
calibration path. The following calibration assumes the scale is set to pounds, and 100 (40 kg) /
200 (80 kg) / 300 (120 kg) / 500 (200 kg) lb is employed as the calibration standards.
Note: Only weights certified and traceable to national standards are to be used for
calibration procedures.
Scale Calibration
Press and hold the LB/KG and ZERO buttons at the same time, press the
ON/OFF button to turn on the scale. "CAL" will briefly appear on the display,
followed by five numeric numbers indicates you can calibrate. Release all
three buttons.
Press the ZERO button. "C-100" (for LB mode) or "C-40" (for KG mode) will
flash on the display, followed by "0".
Place 100 lb or 40 kg on the scale. Five numeric numbers will appear.
When calibration numbers stabilize, press the ZERO button. "C-200" (for LB
mode) or "C-80" (for KG mode) will appear on the display, followed by "0".
Place 200 lb or 80 kg on the scale. Five numeric numbers will appear.
When calibration numbers stabilize, press the ZERO button. "C-300" (for LB
mode) or "C-120" (for KG mode) will appear on the display, followed by "0".
Place 300 lb or 120 kg on the scale. Five numeric numbers will appear.
When calibration numbers stabilize, press the ZERO button. "C-500" (for LB
mode) or "C-200" (for KG mode) will appear on the display, followed by "0".
Place 500 lb or 200 kg on the scale. Five numeric numbers will appear.
When calibration numbers stabilize, press the ZERO button. "END" will
appear on the display, followed by "OFF". The scale will automatically shut
off.
.
ACTION
21
DISPLAY SHOWS
"CAL" followed by
"XXXXX"
"XXXXX" followed
by "C-100" or
"C-40" then "0"
"XXXXX" followed
by "C-200" or
"C-80" then "0"
"XXXXX" followed
by "C-300" or
"C-120" then "0"
"XXXXX" followed
by "C-500" or
"C-200" then "0"
"XXXXX" followed
by "END" then
"OFF"