To Measure DC Current:
Current is measured in amps.
Do not attempt to measure current in circuits capable of
delivering greater than 600V. If the current is unknown,
begin at the highest range, selecting the next lowest until a
reading is displayed.
1. Plug the test leads into the meter inputs as indicated in
the following diagram.
2. Select the proper range to be used within the ADC area.
3. Turn power off.
4. Connect the meter in series with the load or circuit.
5. Turn power on.
6. Measure DC Current.
Meter Setup:
Current <200mA DC
Red
Black
Circuit Connection:
Function
Range
Resolution
DC Current
200.0µA
20.00 mA
200.0mA
20.00A
Page 11
Current >200mA DC
10 Amps Max.
Black
Red
Accuracy
.1µA
± (1.0% +1)
.01 mA
± (1.0% +1)
.1mA
± (1.0% +1)
.01A
± (2.0% +3)
To Use Accessories:
For AC Current Clamp:
1. Plug test leads into the meter inputs as indicated in dia-
gram 1.
2. Remove the probe tips from the end of the leads.
3. Attach the leads to the current clamp (polarity will not affect
reading).
4. Select the mVAC range on the meter.
5. Snap the jaw of the current clamp around one of the current
carrying conductors.
6. Take reading.
For all other Accessories:
1. Plug test leads into the meter inputs as indicated in dia-
gram 2.
2. Remove the probe tips from the end of the leads.
3. Attach the leads to the accessory.
4. Select the mVDC range on the meter.
5. Turn the accessory on.
6. Take reading.