Error type B

Self-test failed
All symbols are flashing.
Repeat the self-test as the CardioMessenger did not
finish it.
The CardioMessenger is not ready for use.
1
Disconnect the CardioMessenger from the
power supply brick if relevant.
2
Press the blue key on the right side of the
CardioMessenger for about two seconds.
3
Leave the CardioMessenger turned off for about
thirty seconds.
4
Turn on the CardioMessenger by connecting it to
the power supply.
The CardioMessenger starts and automatically repeats
the self-test. When the self-test is completed, the
CardioMessenger is ready for use.
Note: The connection test can take up to 15 minutes.
If all symbols continue to flash, the CardioMessenger is
defective. Return it to your physician.