Test of
resis-
tance to
interfer-
ence
Conducted
RF inter-
ferences
according
to
IEC 61000-
4-6
Radiated
RF inter-
ferences
according
to
IEC 61000-
4-3
Test
Com-
level
pliance
accord-
ing to
IEC
60601-
1-2
3 V
3 V
eff
150 kHz
outside
to
the ISM
80 MHz
bands
6 V
inside
the ISM
bands
3 V/m
10 V/m
80 MHz
Up to
to
28 V/m
2.5 GHz
in the
funda-
mental
fre-
quency
bands
of tele-
com-
muni-
cation
equip-
ment
Guidelines for the
electromagnetic envi-
ronment
The minimum distance
eff
of the CardioMessenger
from portable and
mobile radio devices,
including the cables,
eff
should correspond to
the recommended safe
distance that is calcu-
lated according to the
equation for the suitable
transmission frequency.
Recommended safe
distance: d = 1.17 √P
d = 1,17 √P for 80 to
800 MHz
d = 2.34 √P for 800 MHz
to 2.5 GHz with P as the
nominal output of the
transmitter in Watts (W)
according to the infor-
mation from the trans-
mitter manufacturer,
and d as the recom-
mended safe distance in
meters (m).
The field strength of sta-
tionary transmitting
devices should be
measured on site
must be lower than the
compliance level at all
frequencies.