Electromagnetic Emitted Interference According To - BIOTRONIK CardioMessenger Smart Manual Tecnico

Ocultar thumbs Ver también para CardioMessenger Smart:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

34
Appendix
Electromagnetic Emitted Interference
according to IEC 60601-1-2
The CardioMessenger is suitable for operation in the
indicated electromagnetic environment. The customer or
operator of the CardioMessenger should make sure that
it is used in an electromagnetic environment as
described below.
Measurement of
the emitted inter-
ference
RF interference
according to
CISPR 11
RF interference
according to
CISPR 11
Interference of
harmonic oscilla-
tions according to
IEC 61000-3-2
Emitted interference
of voltage fluctua-
tions/flicker
emissions according
to IEC 61000-3-3
Com-
Guidelines for the
pliance
electromagnetic
environment
Group 1 The CardioMessenger
uses RF energy only for
its function. Therefore,
the RF interference
emitted is very low and
not likely to cause any
interference in nearby
electronic equipment.
Class B The CardioMessenger
is suitable for use in all
areas, including living
space and those areas
Class A
that are directly
accord-
connected to a public
ing to
power supply system
IEC 610
that also supplies
00-3-2
buildings intended for
Com-
residential purposes.
plies