Accuracy Table
(6) DC voltage
Measurement value/MAX/MIN/AVE
Range (Auto-range threshold)
600.0 mV
(more than 6000 counts)
6.000 V (more than 6000 counts/
less than 540 counts)
60.00 V (more than 6000 counts/
less than 540 counts)
600.0 V (more than 6000 counts/
less than 540 counts)
1500 V (less than 540 counts)
*1: At DC input
*2: In the 1500 V range, the instrument can withstand input of up to 1000 V continuously or input in excess of
1000 V for no greater than 1 minute.
50
Accuracy guarantee range
(Resolution)
0.0 mV to ±600.0 mV
(0.1 mV)
0.000 V to ±6.000 V
(0.001 V)
0.00 V to ±60.00 V
(0.01 V)
0.0 V to ±600.0 V
(0.1 V)
2
0 V to ±1000 V *
(1 V)
2
±1001 V to ±1700 V *
(1 V)
Measurement
Input impedance*
accuracy
±0.5% rdg.
6.7 M
±0.5 mV
±0.5% rdg.
6.7 M
±0.003 V
±0.5% rdg.
6.1 M
±0.03 V
±0.5% rdg.
6.0 M
±0.3 V
±0.5% rdg.
6.0 M
±3 V
±2.0% rdg.
±5 V
1
Ω
±5%
Ω
±5%
Ω
±5%
Ω
±5%
Ω
±5%