Materials used

- Nebuliser: Polypropylene, thermoplastic
elastomers, polyoxymethylene
- Mouthpiece: Polypropylene, silicone
rubber
®
The eFlow
rapid contains no parts made
from natural rubber (latex).
Ambient conditions for
operation
- Ambient temperature:
+5 °C to +40 °C
- Relative humidity of ambient air:
15% to 93% (non-condensing)
- Air pressure: 700 hPa to 1060 hPa
Within professional care settings, the
device must only be used in in-patient
wards and intensive care units. The
device must never be used in areas
exposed to elevated magnetic or electrical
radiation (e.g. in the vicinity of an MRI
scanner).
Users can operate the device both at
home and when out and about. Its use on
trains and aeroplanes is restricted to pas-
senger areas only. If using the device in a
car, it must be operated with battery
power.
NOTE:
When using portable wireless communi-
cation devices, you must ensure that a
distance of at least 30 cm is maintained
between all parts of the nebuliser sys-
tem and these devices (including their
accessories, e.g. antenna cables or
external antennas). Otherwise, the per-
formance of your nebuliser system may
deteriorate.
50
Storage and transportation
conditions
- Temperature: -25 °C to +70 °C
- Relative humidity of ambient air:
0% to 93% (non-condensing)
- Air pressure: 500 hPa to 1060 hPa
NOTE:
Moisture from condensation can impair
the controller's functional capabilities.
Therefore, you should avoid extreme
temperature fluctuations. Wait until the
controller has reached room tempera-
ture before beginning an inhalation
session.
If the device is not expected to be used for
a prolonged period, the batteries or
rechargeable batteries must be removed.
Info:
Batteries vary considerably in terms
of storage life and output depending
on battery quality. The life cycle of
about 90 minutes can only be
achieved with the recommended
batteries (4x).
®