CLEANING AND DISINFECTION

5
The nebuliser handset (and the aerosol
head) must be cleaned immediately after
each use, and disinfected once a day.

Preparation

Disassemble the nebuliser system into its
constituent parts:
• Pull the connection cord out of the
nebuliser handset:
• Remove the mouthpiece from the
nebuliser.
• Carefully pull the blue expiratory
valve (7a) on the mouthpiece (7) out of
the slot.
Info:
The expiratory valve is held in place
by a thin fixing clip so that it cannot
get lost. This clip must not be
detached from the mouthpiece.
7a
• Open the medication reservoir and pour
out the medication residue.
40
• Disassemble the medication cap.
To do this, press the cap seal (4a)
downwards and out of the cap (4b).
• Open the aerosol chamber.
• Remove the aerosol head.
To do this, gently press the engaging
hooks on the aerosol head together
from the side:
7
• Remove the inspiratory valve from the
aerosol chamber.
4b
4a
®
eFlow
rapid - 01/17