Quick Start Guide
Button
Short Press
Turns the
balance on;
If balance is
On, sets Zero
Long Press
Turns the
balance Off
Short Press
(Menu
Accepts the
Function)
current setting
CALIBRATION
When the balance is operated for the first time, a span calibration is
recommended to ensure accurate weighing results. With balance turned on:
Step-1
Press and hold Menu until [mMeNU] (Menu) is displayed.
Step-2
Release the button, the display will show [C.A.L].
Step-3
Press Yes to accept. [SpaN] will then be shown.
Step-4
Press Yes to begin the span calibration. Zero reading is stored
automatically.
Step-5
Place the specified span calibration mass on the pan. Then, press Yes to
accept the weight or No to select an alternate weight.
[--C--] will be displayed while the reading is stored
The display will show [done] if the calibration was successful. The balance
returns to the previous application mode and is ready for use.
OPERATION & APPLICATIONS
The balance can be configured to operate in various Application modes. Press
and hold Mode key to select your Application mode.
The Scout SPX balance incorporates the following Applications:
[Count]
[wWEIGH]
Weighing
Parts
Counting
Ohaus Corporation www.ohaus.com
Zero/On
Initiates print
Off
Changes the
weighing unit
Yes
Advances to
the next menu
or setting
[Percnt]
[CHECk]
Percent
Check
Weighing
Weighing
Scout
Print
Function
Initiates an
application
mode
Units
Mode
Changes the
application
mode
No
Back
Moves to
previous
menu item
[total]
Totalization
iii
SPX Balances
Tare
Tare
Menu
Enters the
User menu
Exit
Exits the User
menu
[Hold]
Display
Hold
Scout
SPX Balances