Programming Patient Positions - Sirona Dental C2 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

8 Programming Patient Positions

8
Programming Patient Positions
18
Sirona Dental Systems GmbH
Chair and dentist unit
The following four factory-set programs can
be individually reprogrammed by operators
A and B.
(Overwriting of factory default setting).
Programs 1 and 2 (freely assignable)
Neutral program 0.
Mouth rinsing program S.
Move the chair to the desired treatment
position by pressing the corresponding
setting keys on the right-hand keypad
(see page 17).
If you wish that after reaching the pro-
grammed treatment position that the
SIROLUX FANTASTIC unit lamp
should switch on or remain off, then you
must now switch the lamp ON or OFF.
This setting is then also programmed.
Move the dentist unit to the desired
treatment position by hand.
CAUTION: To prevent damage to the
dentist or assistant unit and the
chair upholstery, make sure that nei-
ther unit violates the travel range of
the chair.
To save the program settings, press the
required program key approx. 5 sec-
onds until an acoustic signal sounds.
Programming is completed.
NOTE: The programming can not be per-
formed by activating the program keys
with the foot switch (cursor control).
Faulty programming is avoided in this
way.
D 3311.201.01.10.02
41 63 189 D 3311

Publicidad

Tabla de contenido
loading

Tabla de contenido