The detailed sets are as below:
SERIAL
NUMBER
DISPLAY
Zero
1
Zero tracing
2
DFLOU
Dynamic tracing
3
D-rng
Dynamatic diagnos-
tic
PointD
4
Average value filter
FILtEr
wind-resistance and
5
shock-resistance
filter
Spnt
6
Calibration mode
SPEED
7
Sample speed
8
Z on
7. ATTENTION
1. The balance is precision device, therefore, any pulling and pressing action is prohibited in case that the sensor is deformed or damaged.
2. The balance should be placed on a stable and a smooth flat surface. Strictly prohibit using it in strong air current, strong electromagnetic field
and strong calorific radiation environments.
3. The machine is strictly prohibited from any rain or water. The scale is washable.
4. Any strong corrosive impregnant is strictly prohibited to wash the scale or the circuit board.
5. Don't let the cockroach or parasite get into the machine.
6. If the electronic balance has been moved or when the workplace's temperature fluctuation is significant, the balance should be re-calibrated
to avoid measurement errors.
7. Those easy-volatiles and corrosive product should be put into hermetically sealed containers in case that they would corrode and damage
the balance.
8. The balance is not allowed to overload lest it damage the sensor.
8. WARRANTY
1. The DSN series is warranted for one year from the date of purchase under the terms and conditions of Non-user oriented failure with proper
installment and usage. Please fill out the product and guarantee card and mail the cards to Shangdeli general agency for free repair. (Note: the
sensor and the load cell are not included)
2. If the product is beyond the warranty period or damaged by human factor, we would repair the balance and charge a reasonable fee.
SET
VALUE
SET INTRODUCTION
0.5d
Tout au plus haut c'est la valeur de recherche, plus rapidement celui-ci retournera
1d
à zéro.
2d
1
Suitable for powder and liquid products
4
6
8
Suitable for the sheet ,nubby and granular products
10
0.5d
The higher the dynamic diagnostic value is, the more stable the balance is. But the
1d
resolution becomes bad, so it is better to use in a worse environment.
2d
1
The higher the filter point is, the more stable the scale is.
2
But It responds slowly.
4
0
1
The higher the value is, the more stable the balance is. It is suitable for using under
the worse environment and unsuitable for the powder products.
2
3
0
Single point calibration mode
1
Multipoints calibration mode
0
Fast sampling
1
Factory set
0
1
Factory set
15