10.4.3.3
Dynamic (Animal) Weighing Setup Parameters
When the "
Dynamic
" symbol is displayed, and you have selected "" to set up the
balance for dynamic weighing (see section 10.4.3.3 on Dynamic Weighing Setup
Parameters):
Press [Setup] to select "" to change the dynamic weighing mode settings.
The display will show "". Press [Setup] again and use the [Up] or [Down] keys to
select "" or "".
If "" or "" is selected, the following 4 parameters are available:
Threshold
Mode
Delay
Test time
10.4.3.3.1
Threshold
Press [Setup] when "" is shown and the display will next show the minimum
weight of the item required by the balance to start the process for dynamic weighing. The
value shown will be the current value in the last selected unit.
The minimum threshold value can be changed from 1.0 to 100 grams using the keypad
numeric entry method. If a value outside this range is selected then it will not be accepted
and "" or "" will be displayed followed by return to the weight entry screen
again.
To confirm the desired value, press [Setup] or to escape without changing the value,
press [Mode].
10.4.3.3.2
Mode
Auto "" or Manual "" modes are available. Whichever mode is visible when
[Setup] is pressed becomes the active mode. Auto starts dynamic weighing test as soon
as weight exceeding a set threshold is loaded on the pan. Manual requires the user to
load the pan and then press a button before weighing commences.
10.4.3.3.3
Delay
Press [Setup] when "" is shown and the display will next show the number of
seconds pause before the sampling starts. The Delay value can be changed to between 0-
99 seconds using the keypad numeric entry method. If a value outside this range is
selected then it will not be accepted and "" or "" will be displayed followed
by return to the time entry screen again.
To confirm the desired value, press [Setup] or to escape without changing the value,
press [Mode].
10.4.3.3.4
Test time
Press [Setup] when "" is shown and the display will next show the number of
seconds over which the balance will average to compute the final weight. The Test time
© Adam Equipment Company 2017
""
""
""
""
"" (For Auto mode only)
""
""
""
26