Model #
Maximum Capacity
Readability (d)
Number of intervals
n=
Min. weight
Repeatability
(Std. Dev)
Linearity +
Units of Measure
Stabilization Time
Operating Temp
Power Supply
Input Voltage
Weighing
mechanism
Calibration
External Calibration
Mass
Display
Draft Shield
(w x d x h)
Pan Size
Overall Dimensions
(w x d x h)
Net Weight
© Adam Equipment Company 2017
NBL 623 e / i
620 g
620000
0.02 g
grams, milligrams, carats, grains, Newtons, ounces, troy ounces, pennyweight, custom
15ºC to 35ºC recommended, 40 – 60 % RH (non-condensing)
External mains power adapter - supplied as standard
Suffix i = internal calibration mechanism, e = external calibration only
Recommended OIML class: E2, ASTM / ANSI class: 2
LCD with blue backlight, 7 characters, 20 mm high, and symbols
Glass Ring Draft Shield With Alloy Lid (180 mm diam. x 90 mm)
0.001 g
0.002 g
0.002 g
Typically 3 seconds
(Input Voltage 100–240 VAC, 50/60 Hz)
18 VDC - 830 mA
Force Restoration Balance Motor
500 g
Round, 160 mm diameter
220 x 310 x 90 mm without breeze ring
8.7 x 12.2 x 3.5 in
4.0 kg / 8 lb 13 oz (external calibration model)
4.8 kg / 10 lb 9 oz (internal calibration model)
NBL 823 e / i
820 g
820000
0.02 g
7