Internal parameter table -2
Parameter
Parameter
indication
name
Lc
Password
Temperature
ndA
Alarm mode
Temperature
ndc
control mode
Bit control
dE1
Upper
deviation
Bit control
dE2
Lower
deviation
ndT
Timing mode
Constant
Hn
temperature
timing
Constant
SPd
temperature
deviation
Constant
temperature
SPT
Prompting
time
Timed end
EST
Prompting
time
Whether to
continue the
constant
EH
temperature
control at the
end of the
timing
ndo
Parameter function description
When "Lc=9",the parameter values
can be viewed and modified.
0: Only temperature deviation
overtemperature alarm;
1: At the same time, there are
temperature and lower deviation over
temperature alarm.
0:Fuzzy PID control; 1:Bit control
When the "temperature measurement
> temperature setting value + dE1", turn off
the heating output.
When the "temperature measurement
value is less than the temperature setting
value DE2", the heating output is turned on.
Description: this parameter is effective
only when controlling the position.
0:No timing function;
1:Constant temperature timing;
2:Starting up timing ;
0:Minute time;
1:Hour time ;
When the "temperature measurement
value is more than or equal to the set point
of temperature", it is considered to enter the
constant temperature state.
When entering the constant temperature
state, the buzzer prompts the time.
Note: when "SPT=9999", it represents a
permanent prompt.
When the timing is over, the buzzer prompts
the time.
Note: when "EST=9999", it represents a
permanent prompt.
0:Turn off the heating output after
timing;
1:The constant temperature control is
continued after the timing is finished ;
Reserved, invalid.
10
Factory value
0
(0~1)
0
(0~1)
0
(0~100.0℃)
0
(0~100.0℃)
0
(0~2)
1
(0~1)
0
(0.1~100.0℃)
0.5
(0~9999)
0
(0~9999)
60
(0~1)
0