Internal parameter table -1
Parameter
Parameter name
indication
Lc
Password
Upper deviation
ALH
Over temperature
alarm
Lower deviation
ALL
Over temperature
alarm
P
Proportional band
I
Integration time
d
Differential time
T
Control period
Measurement
Pb
temperature
deviation correction
Measuring
PL
temperature slope
correction
Communication
Addr
address
Loc
Set lock
Parameter function description
The parameter value can be viewed and
modified when "Lc=3".
When "Temperature measurement>
Temperature set value + HAL", there is an
upper deviation over temperature alarm.
When "Temperature measured value
<temperature set value-ALL", there is a lower
deviation over temperature alarm.
Note: When "ALL=0", the lower deviation
alarm is invalid.
Time proportional effect adjustment.
Integral function adjustment.
Differential regulation.
Heating control cycle.
It is commonly used to correct errors that occur
during low temperature measurements.
Pb = actual temperature value - meter
measurement
It is often used to correct errors that occur
during high temperature measurements.
PL = 1000 * (actual temperature value -
meter measurement) / meter measurement
This machine communication address.
0: Can modify the temperature or time
setting value;
1: It is forbidden to modify the
temperature or time setting value.
9
(Range) Default
Value
0
(0~100.0℃)
5.0
(0~100.0℃)
0
(0.1~300.0℃)
10.0
(1~2000 Second)
200
(0~1000 Second)
200
(1~30 Second)
5
(-50.0~50.0℃)
0
(-999~999)
0
(1~32) 1
(0~1) 0