Troubleshooting

The following might indicate an issue with the device:
● excessive fuel gas consumption in comparison with normal
consumption
Fuel gas
40% H
100% H
● constantly fluctuating zero point (up to 3 ppm)
If you believe there is a fault:
1. Close the valve on the fuel gas bottle immediately.
2. Switch off the device.
3. Have the device checked at once by Sewerin Service.
8

Technical data

Dimensions (W x D x H)
Weight
Material
Design
Gas input
(connection to fuel gas bottle)
Gas outlet
(connection to pressure hose)
Inlet pressure
Operating pressure, preset
Setting range
Suitability
, 60% N
2
2
2
Normal consumption
approx. 2 l/h
approx. 1 l/h
110 x 140 x 105 mm
700 g
stainless steel, high-strength aluminium
2-step with inlet pressure compensation
manual connection according to
DIN 477 no. 1
W 21.8 x 1/14 LH for hydrogen
90° angled connection
G3/8-AG
max. 200 bar
1 bar
0 – 1.5 bar
for high-purity gas up to 6.0
English │ 15