Transcend 365 miniCPAP Auto
Humidifier LCD indicators in Standby
Mode
Water Status Indicators
These indicators will only be displayed if humidifier is set to a level > 0. They will appear
in lower right portion of the Home Screen
Flashing Blue Droplet - Out of Water
Indicated by a flashing blue droplet. Humidifier will be disabled during therapy. Can be
caused by:
No water in the reservoir – only displayed if heater/humidifier is set to level > 0.
•
If this status appears but there is water in the reservoir, it indicates an error likely
•
due to moisture in the CPAP. Allowing the system to dry out may correct the issue.
To do this, run the CPAP in Dry Mode. Note: After drying the system, disconnect
and reconnect the power source to ensure the status is reset.