After Use - gce mediline Varimed Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 8
pressure relief valve vent holes.
Turn off the cylinder valve by turning the hand wheel in clockwise direction to
„stop" position. Do not use excessive force
If any leakage is detected, use the procedure in chapter 6.3 and return
the product for service.
6.1.4. Functional test before use
Open the cylinder valve – position "ON".
Check that the pressure gauge indicates pressure.
Turn off the cylinder shut-off valve. Do not use excessive force.
To test the regulator functionality, check gas flow through the regulator. For
regulators with a quick coupler outlet, connect and disconnect the male QC
probe. For regulators with threaded outlet, gas flows already in opening the
cylinder valve.
6.2. USE
Connect required equipment downstream of the regulator ensuring that all
valves are closed.
Open the cylinder valve slowly and check that there are no leaks at the
connections using a proprietary leak detection fluid.
Assurez-vous par la pression indiquée sur le manomètre du détendeur qu'il
y a suffisamment de gaz disponible. Une fois que l'aiguille du manomètre est
stable, ouvrez le robinet de la bouteille en faisant un tour complet.
Sudden opening of the cylinder valve could result in a danger of fire or
explosion arising from oxygen pressure shocks. Insufficient opening of
the cylinder valve could reduce actual flow delivered.

6.3. AFTER USE

Turn off the cylinder valve by turning the hand wheel in clockwise direction to
"stop" position. Do not use excessive force.
Vent the regulator.
Ensure the pressure indicator does not show any residual pressure.
Disconnect all connected equipment from the user outlets.
Never leave regulators under pressure unless they are in use!
CLEANING
7.
Remove general contamination with a soft cloth damped in oil free oxygen
compatible soapy water & rinse with clean water.
Disinfection can be carried out with an alcohol-based solution (spray or wipes).
If other cleaning solutions are used, check that they are not abrasive and that
they are compatible the product materials (including labels) and gas.
Do not use cleaning solutions containing ammonia!
Do not immerse in water or any liquid.
Do not expose to high temperature (such as autoclave).
5/44