SET UP MODE
The set up mode is used • to set the alarm level,
This set up mode may only be entered by healthcare professionals and authorised personnel trained to use this applica-
tion of the pump.
The patient or his/her relatives without permission of the physician, dietician, nurse or other licensed practitioner, may
not change the settings of the set up mode.
+
+
&
–
2 sec
2 sec
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode.
First the alarm level can be set: "BEEP HIGH" (+) or "BEEP LOW" (-)
Use the "+" or "-" key to change the setting from "HIGH" to "LOW" or from "LOW" to "HIGH".
+
LOCK
fill set
–
UNLK
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the
"FILL SET" key to switch to the next setting:
"UNLK" the keypad (-) or "LOCK" the keypad (+). In the "LOCK" mode the feeding program of the pump
can no longer be changed.
Use the "+" or "-" key to change the setting from "LOCK" to "UNLK" or vice versa.
+
BEEP WHEN DONE
fill set
–
MUTE WHEN DONE
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the
"FILL SET" key to switch to the next setting:
"BEEP WHEN DONE" (+): the pump will give an alarm when the dose is administered or "MUTE WHEN
DONE" (-): the pump will not give an alarm after administering the dose. Use the "+" or "-" key to change
the setting from "BEEP WHEN DONE" to "MUTE WHEN DONE" or vice versa.
+
LITE ON
fill set
–
LITE OFF
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the
"FILL SET" key to switch to the next setting:
ZMNL900617_AW.indd 88
• to lock the keypad,
• to switch the audible alarm off when a dose is administered.
• to switch the light of the LCD permanently on when connected to the mains.
• to enable connection with a Nurse call or Patient Data Monitoring System
(PDMS).
TO SET THE ALARM LEVEL
BEEP HIGH
BEEP LOW
TO LOCK THE KEYPAD
TO MUTE WHEN DOSE DONE
TO SWITCH LIGHT ON
-88-