Routine Maintenance; Nimbus 4 And Nimbus Professional Systems; Nimbus Pump; Nimbus 4 And Nimbus Professional Mattresses - Arjohuntleigh Nimbus 4 Instructiones De Uso

Ocultar thumbs Ver también para Nimbus 4:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 119

8. Routine Maintenance

Nimbus 4 and Nimbus Professional Systems

Maintenance
Servicing
Service Period

Nimbus Pump

General Care,
Maintenance and
Inspection
Biofilter

Nimbus 4 and Nimbus Professional Mattresses

General Care

Serial Number Labels

Mattress
The equipment has been designed to be virtually
maintenance-free between service periods.
ArjoHuntleigh will make available on request service
manuals, component parts lists and other information
necessary for ArjoHuntleigh trained personnel to repair
the system.
ArjoHuntleigh recommend that the Nimbus 4 and
Nimbus Professional systems should be serviced after 12
months continuous running time, by an ArjoHuntleigh
authorised service agent. This is indicated by the
illumination of the Service symbol (refer to "Service
Indicator" on page 12).
Check all electrical connections and the mains power
cable for signs of wear or damage.
Test the Power Fail Alarm before use (refer to "Testing
the Power Fail Alarm" on page 8).
If the pump has been subjected to abnormal treatment,
e.g. immersed in water or dropped, the unit must be
returned to an authorised service centre.
The internal biofilter can be run continuously for two
years before it requires autoclaving or replacement.
The biofilter can only be replaced by a service engineer.
Remove the cover from the mattress.
Inspect the cover for signs of wear or any tears, and
check that all cover fasteners are secure.
Check the security of all internal connections, including:
• Between the cells and the manifold.
• To the CPR/Transport Control.
Make sure all cell fasteners are correctly connected to
the mattress base sheet and are not loose or damaged.
Make sure the drag handles are fastened to the base
cover to remove the risk of tripping.
The serial number label is on the back of the pump case.
Pump
The serial number label is on the top of the CPR/
Transport Control.
Quote these serial numbers when requesting service.
23

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Nimbus professional

Tabla de contenido