Control Dial Settings - Abbott Patrol Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

CONTROL DIAL SETTINGS

DIAL:
OFF/
CHARGE
SET RATE
SET DOSE
RUN
FUNCTION:
Stops all pump functions. Battery charges in all dial
settings when pump is plugged in. The pump will
retain current RATE, DOSE and VOL FED values
indefinitely unless the pump is unplugged from an AC
power source while in the OFF/CHARGE setting.
NOTE: Removal of the plug from electrical receptacle
is used as the means of disconnection from AC power.
Allows flow rate to be set (1 to 300 mL/hr in
1-mL/hr increments) by pressing the keypad arrows
when the pump is in the SET RATE mode. The up
arrow increases the rate. The down arrow decreases
the rate.
NOTE: The LOC feature may be activated by
depressing both arrows simultaneously when the dial
is in the SET RATE position. Repeat procedure to
deactivate. The SET RATE, SET DOSE and CLEAR
VOL values cannot be changed while the LOC feature
is activated.
Allows dose to be limited (1 to 9999 mL in 1-mL
increments) by pressing the keypad arrows when
the pump is in the SET DOSE mode. The up arrow
increases the dose. The down arrow decreases the
dose. Completion of this dose triggers the doSE FEd
alarm. Refer to doSE FEd (page 12).
NOTE: Pump will operate without a DOSE selected
(SET DOSE at 0000).
Only setting that activates feeding. Visual display shows
rate of flow in mL/hr and the dash on the left side of the
display scrolls from top to bottom when pump is in RUN
mode.
NOTE: With flow rates less than 185 mL/hr, it is
normal for rotor movement to be intermittent (e.g.
starts, stops, starts, stops, etc) while in the RUN mode.
This intermittent movement regulates the rate of flow,
conserves battery life and does not affect the accuracy of
the pump. The RUN mode should always be verified by
the top-to-bottom scrolling bars on the left side of the
display window.
Battery Operation Indication: Battery-powered operation
is indicated by periodic flashing of the displayed SET
RATE while in the RUN mode. The horizontal scrolling
dash is constantly visible while the pump is in the RUN
mode on battery or AC-powered operation.
6

Publicidad

Tabla de contenido
loading