Specifications - Abbott Patrol Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

SPECIFICATIONS

POWER
Voltage:
Power:
Fuse:
Line Cord:
Leakage:
MECHANICAL
Height:
Width:
Depth:
Weight:
OPERATIONAL SPECIFICATIONS — FLOW RATES
Range:
Increments:
Accuracy:
Pressure:
Operating Temperature: 50˚F - 104˚F (10˚C- 40˚C)
Operating Humidity:
Operating Altitude:
STORAGE
• Be sure pump dial is in the OFF position.
Storage Temperature:
Storage Humidity:
Storage Altitude:
100 to 130 VAC, 60 Hz, 1-Phase
15 watts, average
200 mA, time-delay type
Hospital-grade, nondetachable (10 feet)
Less than 100 microamps
8.5 inches
6.5 inches
4.8 inches
6.6 pounds
1 - 300 mL/hr
1 mL/hr
±10% or ±0.5 mL/hr, whichever is greater (refer to
CHECKING PUMP ACCURACY, page 12)
Pumps against 18 psi (124 kPa) nominal back pressure
before no FLO alarm.
20% RH - 80% RH, non-condensing
The system is capable of operating as specified at
elevations up to 10,000 feet (3,050 meters) above sea
level.
32˚F - 95˚F (0˚C - 35˚C)
10% RH - 90% RH, non-condensing
The system is able to withstand and operate after exposure
to elevations up to 20,000 feet (6,100 meters) above sea
level.
17

Publicidad

Tabla de contenido
loading

Tabla de contenido