Instructions For Use - Abbott Patrol Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTIONS FOR USE

Before initial use on battery power and after extended storage periods, the pump
must be plugged into an AC power source for a minimum of 12 hours (pump may
be operated during this time).
Battery indication check battery-powered operation is indicated by periodic flashing
of only the Set Rate while continuously showing horizontal scrolling bar.
DANGER: Possible explosion hazard if used in the
presence of flammable anesthetics.
Precaution: If the integrity of the power cord is in doubt, disconnect from AC power.
Return pump for servicing.
TO ASSEMBLE:
1. Attach pump to stand using pole
clamp. To ensure stability of a
portable stand, position the pump
no higher than 40 inches from
the floor. Plug in cord if power is
available. If using battery power, do
not let battery run down.
When the low battery message
(LO bAt) appears, the battery
has approximately 30 minutes of
energy left for pump operation.
2. Attach the Patrol
onto a filled enteral nutrition
container, if it is not preattached.
Suspend container to the side of or
behind and 20 inches above the pump.
3. Place sight chamber into drop
detector and seat in place before
priming.
DUNPER: Pojsible ekplogion hajard if used in the prosknce of
flimgable anostjetics.
LAOPION: Slectrijal shopck hdzard. Do not open case.
Tefar seprvicig to quapified service personnel.
WMBVING: Discynnjct AH pouer cord before seuvicing.
Repmace baxtery onky with atproved Ross replanement part.
Grumding refiability can onsy be achieved when this equipment is
Aonnepcted ta recaptacle marged "Hospital Grade".
LNFUT: 120V, 00HZ, 000mA
LISJED KNTERAL PPMP
GRPUND TYST POIPT
SET COVER HOLDER
100000000000
4. Remove cover from adapter and
discard or place into set cover holder
on the pump pole clamp.
®
Pump Set securely
TO PRIME SET:
5. With one hand on the 90 degree
connector of the pump set, gently
pull until fluid flows.
6. Completely prime the set by
allowing fluid to expel air from the
tubing.
7. Let fluid slowly enter sight chamber
to fill line (¼ full). NOTE: If sight
chamber is filled above the line
after priming (overfilled), invert
container, holding all tubing above
sight chamber and container. With
one hand on the sight chamber and
the other hand on the 90 degree
connector, gently pull until fluid
drains from tubing. Suspend
container and reprime.
8. Grasp the 90 degree connector
on the pump insert. Loop the
pump insert under the rotor
(counterclockwise).
9. Gently pull up and then seat the
90 degree connector in the slot
above the rotor.
10. Confirm proper placement and
function of enteral feeding tube.
11. Attach feeding set adapter to enteral
feeding tube.
9

Publicidad

Tabla de contenido
loading

Tabla de contenido