TESMED MAX 830 TECHNICAL DATA

Pulse repetition frequency
Max energy per pulse
Max peak power / mean power
Mean current at max frequency
Mean current at min frequency
Max peak current
Pulse duration (Positive/Negative)
Pre-set programs
Number of channels
Dimensions (mm)
Applied part: Low Voltage
Conditions of use
All electrical data are provided for a charge of 1000 ohm per channel
Equipment not suitable for use in the presence of inflammable anaesthetic mixtures with air or oxygen or nitrous oxide.

BATTERY CHARGER TECHNICAL DATA

Electrical output
Polarity, external Ø, internal Ø
Compliance with technical standards
Power supply
Casing
Pulse shape
Weight (gr)
Electrical input
Nickel-metal-hydrate (NI-MH)
rechargeable battery 4.8V = 750 mA/h
10 W per pulse / 5 mW
(positive and negative),
no continuous component
100-240V∿, 50/60Hz, 180mA
- 36 -
Polycarbonate
1Hz - 150Hz
E= 3200 μJ
2 mA
0,2 mA
120mA
Rectangular, biphase
da 40 μs a 400μs
99
4
128 x 58 x 18
145
BF
Continuous use
6V⎓, 60mA
3mm, 1.1mm
EN 60601-1, 3a ed.