PROBLEM
POSSIBLE CAUSES
Intermittent output
Lead wire
Improper electrode placement.
Stimulation is ineffective.
Unknown
Using electrodes on the same site every time.
The skin becomes
Electrodes are not adhered to the skin properly.
red and/or you feel
a stabbing pain.
The electrodes are dirty.
The surface of the electrode is scratched.
The electrodes come off the skin.
Output current stops
during therapy.
The lead wire is disconnected.
The batteries' power has been exhausted.
EN
POSSIBLE SOLUTION
Verify connection is secure. Insert wire firmly.
Turn down the intensity. Rotate lead wire in socket 90°. If still intermittent,
replace lead wire.
If still intermittent after replacing lead wire, a component may have failed. Call
the repair department.
Reposition electrode.
Contact clinician.
Reposition the electrodes. If at any time you feel pain or discomfort, stop use
immediately.
Ensure the electrodes are securely adhered to the skin.
Clean the electrodes according to description in this manual or replace with
new electrodes.
Replace with new electrodes.
Turn off the device and place the electrodes on again, or replace with new
electrodes.
Turn off the device and connect the lead wire.
Replace with new batteries.