TABLE 2:
GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of these electrical stimulators should assure that it is used in such environment.
Immunity test
Electrostatic discharge (ESD)
lEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips,
short interruptions and voltage
variations on power supply
IEC 61000-4-11
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
EN
IEC 60601
Compliance level
test level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
3 A/m
3 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Not applicable
Not applicable
Not applicable
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.