OPL-ADVANTAGE – OPL system
3.2.14 Set the lock-out time
The lockout feature provides the option of preventing unwanted injections when supply signals are received
more than once in a wash formula. (Not recommended in drain mode).
The lockout available can be set for 0 – 75 minutes commencing when a pump stops. Additional signals for
that pump are ignored for the set lock-out time. If the load count pump receives a signal during the lock-out
period, the lock-out will terminate and prepare to accept the next formula.
Example: If you set a 1 minute lock-out time, after the end of dosing of one pump, the OPL-
ADVANTAGE will ignore the activation of the same signal for 1 minute.
The programming procedure is the following:
1) Enter access code (see par. 3.2.1) ;
2) Select PUMP "2" and FORMULA "A";
3) Use
to set the time.
4) Press ENTER;
5) The display will be flashing briefly to indicate that the new status has been accepted.
6) Press RESET.
In a lock-out is used in drain mode we suggest a lock-out time longer than the sum of the
dosing and the delay time of the pump.
3.2.15 Level enable/disable:
In normal operation, choose this option to disable the second level injection. The pumps will accept signals
and run as normal with any future signals from the washer. (If the pump is not in a lockout time) In situations
where the load count signal is skipped, either from operator error or machine malfunction, disabling levels
can prevent a missed injection. This feature should be used in normal operation mode only.
The programming procedure is the following:
1) Enter access code (see par. 3.2.1) ;
2) Select PUMP "7" and FORMULA "A";
3) Use
to select 000 "Level enable" or 001 "Level disable"
4) Press ENTER;
5) The display will be flash briefly indicating the enable/disable status is set.
6) Press RESET.
OPL-ADVANTAGE
Enable
code 0000136607
Disable
rev. 1.6 EN
13