5. Default settings
To reset the default settings, press the RESET button using a pointed object.
X. TECHNICAL SPECIFICATIONS AND PRECISION
1. Normal use conditions : Operating temperature: 10°C~40°C (50~104°F). Nor-
mal conditions of storage and transportation temperature : -20°C~50°C (-4°F~
122°F). Humidity level: 10 - 85%
2. Measurement display unit: °C or °F.
3. Display resolution: 0.1°C (0.1°F).
4. Measurement range:
In BODY mode: 32.9°C - 42.9°C (90°F – 109°F).
In SURFACE TEMP mode: 0°C - 60°C (32°F - 140°F).
5. Consumption: ≤ 50mW.
6. Measurement distance: 5 cm (2 in).
7. Automatic switch-off: 5 seconds.
8. Accuracy: 36°C - 39°C (96.8°F - 102.2°F) = ± 0.2°C.
9. Battery voltage: DC 3.7V.
10. Charging time: 3 hours
11. Use of battery at full charge: > 10,000 cycles.
12. Battery life: > 3,600,000 cycles.
13. Calibration check: 100,000 cycles.
14. Net weight: 180g.
15. Size: 155 x 134 x 66 mm.
Recharging base station
1. Input: Voltage AC 100~240V – 50~60Hz.
2. Output: DC 6.0V 400mA - Power cable VDE 2 plug.
The THERMOFLASH® PRO LX-261 EVOLUTION can take temperature measure-
ments below 32.9°C or above 42.9°C. Outside of these temperature ranges, preci-
sion cannot be guaranteed.
Compliance
This device meets the requirements of EMC Standard CEI 60601-1-2 Edition 2 con-
cerning medical electrical devices. The manufacturer shall not be held liable for any
interference caused by accessories other than those supplied or by unauthorized
changes or modifications made to the device. Any unauthorized changes or modifi-
cations made by the user may result in the user forgoing the right to use the device.
28
Visiomed - PRO LX-261E - UK - 122015
XI. CLEANING INSTRUCTIONS
- The protective glass over the lens is the most important and fragile part of the
thermometer, please take great care of it.
- Clean the glass with cotton fabric, wet with water or 70° alcohol.
- Do not immerse the device in water to avoid damage.
- Disconnect the power source from the device and remove the battery before
cleaning.
- Clean all parts after use as indicated in the manual.
- Ensure that the electrical connector socket does not come into contact with the
water solution or disinfectant (socket at the base of the device).
XII. ACCESSORIES SUPPLIED
Charger and recharging base station, Wrist strap, Double-sided adhesive patch,
Wall-fixing screws, User manual, Guarantee card.
XIII. TROUBLESHOOTING
If you have one of the following problems while using your THERMOFLASH® PRO
LX-261 EVOLUTION, please refer to this breakdown service guide to help resolve
the problem. If the problem persists, please contact our customer service at +33
892 350 334.
- The screen displays temperature superior to 95°: the temperature is in Fah-
renheit. Change the measurement unit to Celsius.
- The screen displays the body temperature inferior to 32°C (89.6°F) : to measure
a body temperature, the BODY mode must be selected. If you are on SURFACE TEMP
mode, the 32°C (89.6°F) temperature displayed is showing the external temperature
that your body releases.
- The screen displays the message HI : when using the THERMOFLASH® PRO LX-
261 EVOLUTION, the message HI can show on the screen. The analysis is above
the measurement range selected, either superior to 42.9°C (109°F) in BODY Mode
or superior to 60°C (140°F) in SURFACE TEMP Mode.
- The screen displays the message LO : when using the THERMOFLASH® PRO
LX-261 EVOLUTION, the message Lo can show on the screen. The temperature
analysed is under the measuring range selected, either less than 32°C (90°F) in
BODY Mode or less than 0°C (32°F) in SURFACE TEMP Mode.
Visiomed - PRO LX-261E - UK - 122015
29