1. Working according to manual will enable to
obtain expected accuracy and recurrence in
moisture measurement.
It is recommended to make three measurements.
The avearge of three measurements is the final
result.
2. Temperatures of sample and instrument
should be similar (sensor reacts with some
delay).
3. Cleanness of sample for measurement has
an influence on obtaining accurate result.
Sample should not contain husks, tailings
and dirtinesses.
4. You should check if there are no seeds in meas-
urement chamber before work. You should re-
move
any
remainings
measurement from the chamber. Before turn-
ing on an instrument measurement cham-
ber should be clean and empty.
36
CHAPTER 2
PRELIMINARY REMARKS
5. The way of filling the measurement chamber
6. Producer suggests using alcaline batteries. The
7. When the limit ranges of measurement are ex-
"<8%" (below range), when moisture of meas-
ured sample is lower than 8%.
">35,0%" (above range), when moisture of
measured sample is higher than 35%.
8. If important variations of readings appear in
from
earlier
is important and may influence precision of
measurement results.
electrolite which comes out of ordinary battery
after exhausting may damage the device. Al-
caline battery works longer and there is no risk
of outflow of electrolite.
ceeded the display shows:
comparison with laboratory results of the sam-
ples, you should use modification mode (see
the chapter 4: "DATA MODIFICATION").