Operational Cautions - SUEZ Sievers M9 Serie Manual De Operación Y Mantenimiento

Analizadores de toc
Ocultar thumbs Ver también para Sievers M9 Serie:
Tabla de contenido

Publicidad

The UV lamp and the display screen contain mercury and may be considered
hazardous material in your local area. Dispose of these items in accordance
with federal, state, or local government regulations.
In the case of any broken or damaged UV lamp, handle the remains in
accordance with your organization's toxic waste handling procedure and
dispose of them in accordance with federal, state, or local government
regulations.
This symbol indicates the European Union Directive 2002/96/EC for
waste electrical and electronic equipment (WEEE) requires disposing
the Analyzer separately from standard waste.
O
C
PERATIONAL
AUTIONS
To ensure optimal sampling results and to protect the instrument from any damage,
review and implement the following Caution statements.
Make sure the DI water Reservoir is full, particularly when running samples
with high TOC or high salt concentrations. After running high TOC or salt
samples, always rinse the Analyzer by running low-TOC DI water through a
sample cycle.
To avoid false TOC readings and possible damage to the Analyzer, always
make sure the sample inlet is open and the DI water reservoir is filled before
starting analysis
Operation of the Analyzer without the in-line filter on the sample inlet line will
damage the Analyzer and void the warranty. To avoid damaging the Analyzer,
install the filter and replace the filter element, as needed.
To avoid damage to the DI water pump, ensure that the DI water reservoir is
filled BEFORE turning O
To avoid false TOC readings and possible damage to the Analyzer, always
make sure the sample is flowing through the iOS System and the DI water
reservoir is filled before starting analysis.
For performance within specifications on ozonated water systems, an Ozone
Destruct Kit must be purchased from SUEZ and installed according to
instructions.
Analizadores de TOC Sievers M9 y M9
DLM 77000-08 ES Rev. A
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
H
M
AZARDOUS
C
AUTIONS
the Analyzer.
N
e
327 / 418
D
ATERIAL
ISPOSAL
Manual de Operación y Mantenimiento
A
DVERTENCIAS
SUEZ © 2018