Code 66 screen
I-neb is being held at the wrong angle.
Solution: Hold the I-neb horizontally with
the screen facing down. Continue to breathe
normally through the I-neb (see step 16).
Blank screen
Dead battery.
Solution: The I-neb can be used when
connected to the battery charger. To continue
with your treatment, attach the I-neb to the
battery charger and plug into an electrical
outlet (see step 2). Switch I-neb on, wait for
the start screen to appear. Pick up the I-neb
and start your treatment.
Once your treatment is complete, recharge the
battery (see step 2)
Start up screen
Your drug dose (25 or 50) should appear.
18
12035_RRDD_InebPH_InstESUS_1056844_Issue1_v10.indd 20
Treatment in progress
The mouthpiece vibrates when the medication
is delivered.
Treatment successfully completed
Pause Mode (screen flashing)
Solution: To continue, you must start breathing
within 10 minutes. Pick up the I-neb and start
breathing normally. It will take a few breaths
for the I-neb to start delivering medication.
Any other screen display? Contact patient
support.