Guidance and Manufacturer's Declaration – Emissions All Equipment and
Systems
This device is intended for use in the electromagnetic environment specified below. The
customer or user of this device should assure that it is used in such an environment.
Immunity Test
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
Electrical Fast
Transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
SE-DV54-1
32
IEC 60601 Test
Compliance
Level
Level
±6kV contact
±6kV contact
±8kV air
±8kV air
±2kV on AC
±2kV on AC
Mains
Mains
±1kV
±1kV
Differential
Differential
±2kV
±2kV
Common
Common
>95% Dip for
>95% Dip for
0.5 Cycle
0.5 Cycle
60% Dip for
60% Dip for
5 Cycles
5 Cycles
30% Dip for
30% Dip for
25 Cycles
25 Cycles
>95% Dip for
>95% Dip for
5 Seconds
5 Seconds
DEclaration
Electromagnetic Environment -
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are synthetic,
the relative humidity should be at
least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of this
device requires continued operation
during power mains interruptions, it
is recommended that the device be
powered from an uninterruptible
power supply or battery.