Returns; Material Loading; General; Process Connections - ACS contsys Flowcont M Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Information! In accordance with EU guideline governing hazardous materials
Anyone in possession of special-category waste is responsible for its disposal and/or must note the fol-
lowing rulings if transporting such waste:
!
D184B112U04
Undertake regular servicing to check:
– the pressure-resistant panels/liner material of the pressure equipment
– the function
– the seal integrity
– the wear level (corrosion, abrasion, cavitation)
1.1.15

Returns

If you wish to send the equipment to the ACS parent company in Eggenfelden for repairs or re-calibra-
tion, use the original packaging or a suitable secure transport container. Please specify the reason
for returning the equipment.
All flow recorders and/or flow measurement transformers delivered to ACS for repairs must not con-
tain any traces of hazardous substances (acids, lyes, solvents, etc.). In order to do this, hazardous
substances should be rinsed out of the measurement pipe and neutralised. These measures should
be confirmed in writing in the accompanying papers.
If anyone in possession of these hazardous substances is not able to completely remove them from,
the equipment should be transported with the necessary accompanying papers. The owner of the
equipment will be invoiced for any costs which ACS may incur through the disposal of hazardous
substances during repairs.
1.2

Material loading

1.2.1

General

Attention!
Limitations in the equipment's permissible fluid temperature (TS) and permissible pressure (PS) result
from the liner and process connection material used as well as the Viton gasket used. Refer to factory
panel and type plate on equipment.
1.2.2

Process connections

Material of loaded components:
PEEK (DN 10, 15, 25); PVDF (DN 50)
Operating data
Nominal pressure : PN10 ( for max. perm. levels of operating pressure as function of temperature,
refer to p/T rating)
Max. operating temperature : 110°C
Fig. 1:
Flowcont M
p/T rating for Miniflow DN10 to 50
10
9
DN 10
8
7
6
5
4
3
2
-10
0
10 20 30 40 50 60 70 80 90 100 110
max. permissible fluid temperature [˚C]
1 Safety instructions
DN 15 to 50
GB 7