Hart 5437 Manual Del Producto página 54

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Incident Description
An exception error occurred in the
main CPU program execution
An unrecoverable memory error
occurred in the internal main CPU
An unrecoverable memory error
occurred in the internal main CPU
An unrecoverable memory error
occurred in the internal main CPU
An unrecoverable memory error
occurred in the internal input CPU
A critical measurement error is
detected on internal voltage reference
A critical measurement error is
detected on internal voltage reference
A critical measurement error is
detected on Input 1
A critical measurement error is
detected on Input 2
A critical measurement error is
detected on the ground measurement
The device is in simulation mode and
one or more of its Device Variables are
not representative of the process
54
Description
LED reaction
Stack integrity error
Lights Red
CRC16 error in
Lights Red
factory data in
FLASH
RAM cell error
Lights Red
Safe domain RAM
Lights Red
integrity error
CRC16 error in input
Lights Red
CPU configuration
Drift error, reference
Flashing Red
voltage FVR
Drift error, reference
Flashing Red
voltage VREF
Drift error, primary
Flashing Red
Input 1
Drift error, primary
Flashing Red
Input 2
Drift error, ground
Flashing Red
voltage offset to
terminal 3
Device Variable
No Impact
Simulation Active
5437V103-ES
Analog Output
NE-107 Class
Reaction
Safe State
Failure
Safe State
Failure
Safe State
Failure
Safe State
Failure
Safe State
Failure
Safe State
Failure
Safe State
Failure
Safe State
Failure
Safe State
Failure
Safe State
Failure
No impact
N.A.
User action
Reset or re-power the device.
If the error is persistant send in the
device for repair
Reset or re-power the device.
If the error is persistant send in the
device for repair
Reset or re-power the device.
If the error is persistant send in the
device for repair
Reset or re-power the device.
If the error is persistant send in the
device for repair
Reset or re-power the device.
If the error is persistant send in the
device for repair
Reconnect or repair sensor.
If the error is persistant send in the
device for repair
Reconnect or repair sensor.
If the error is persistant send in the
device for repair
Reconnect or repair sensor.
If the error is persistant send in the
device for repair
Reconnect or repair sensor.
If the error is persistant send in the
device for repair
Reconnect or repair sensor.
If the error is persistant send in the
device for repair
N.A.
Error
#
36
37
38
39
40
41
42
43
44
45
46

Publicidad

Tabla de contenido
loading

Productos relacionados para Hart 5437

Tabla de contenido