Main Performance; Main Parameters - Mobiclinic PX-03 Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL FOR PULSE OXIMETER/
MANUAL DE USO PARA PULSIOXÍMETRO
Model/Modelo: PX-03
drug, the SpO
waveform (PLETH) will decrease. In this case, the measurement
2
will be more sensitive to interference.
2. For those with a substantial amount of staining dilution drug (such as methylene
blue, indigo green and acid indigo blue), or carbon monoxide hemoglobin
(COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin, and some with
icterus problem, the SpO
3. The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also
be a major factor blamed for serious error of SpO
4. As the SpO
value serves as a reference value for judgement of anemic anoxia
2
and toxic anoxia, some patients with serious anemia may also report good SpO
measurement.
4 Technical specifications

4.1 Main Performance

 SpO
value display
2
 Pulse rate value display, bar graph display
 Low-power indication: low-power indication symbol appears before working
abnormally which is due to low-power.
 Pulse sound indication
 With alarm function

4.2 Main Parameters

A Measurement of SpO
Measuring range: 0%~100%
Accuracy: 70~100%: ±2%; Below 70%: unspecified.
B Measurement of pulse rate
Measuring range: 30bpm~250bpm
Accuracy: ±2 bpm or ±2% (select larger)
C Resolution
SpO
: 1%, Pulse rate: 1bpm
2
D Measurement Performance in Weak Filling Condition:
SpO
and pulse rate can be shown correctly when pulse-filling ratio is 0.4%
2
SpO
error is ±4%; pulse rate error is ±2 bpm or ±2% (select larger)
2
determination by this monitor may be inaccurate.
2
2
measure.
2
2

Publicidad

Tabla de contenido
loading

Tabla de contenido