b. In the close state of alarm, press button to turn on/off pulse sound.
c. In the alarm state, press button to suspend alarm, and alarm function will be
renewed in about 30 seconds. If press button again in 30 seconds, the pulse sound
will be turned on/off.
If low-power alarm occurs, please charge the battery; and if the battery has
been out of normal life, please replace it by a new one.
7.8 Charge
Put the rechargeable buttony lithium battery properly according to the polarity
direction in the charger. Connect one end of the adapter to the power supply socket,
and connect another end of the adapter to charger with data line. In the charge
process, the indication light of the charger is red. After the battery is full, the
indication light of the charger becomes green.
The power adaptor of the device accords with the requirements of IEC 60601-1.
The charger accords with the requirements of IEC 60601-1.
In charging state, the red indication light of the charger will shine, the green
indication light of the charger shining means the charge has been accomplished.
8 Repairing and Maintenance
Please change the battery when the low-voltage displayed on the screen.
Please clean the surface of the device before using. Wipe the device with
medical alcohol first, and then let it dry in air or clean it by dry clean fabric.
Using the medical alcohol to disinfect the product after use, prevent from cross
infection for next time use.
Please take out the battery if the oximeter is not in use for a long time.
The best storage environment of the device is - 40ºC to 60ºC ambient
temperature and not higher than 95% relative humidity.
The device needs to be calibrated once a year (or according to the calibrating
program of hospital). It also can be performed at the state-appointed agent or just
contact us for calibration.
High-pressure sterilization cannot be used on the device.
Do not immerse the device in liquid.
It is recommended that the device should be kept in a dry environment.
Humidity may reduce the useful life of the device, or even damage it.